1. Device Overview
The Daavlin 4 Series home phototherapy device uses the ClearLink™ Controller and supports three modes of operation: Guided, Dosimetric, and Timed. The controller records treatments and can limit exposures via FlexRx® under physician supervision.
2. Indications for Use
For treatment of diagnosed photoresponsive skin disorders (e.g., psoriasis, vitiligo, atopic dermatitis) under a physician’s direction. Usable from pediatric (with responsible adult) to geriatric populations.
3. Warnings & Safety
- Prescription Use Only. Always follow physician instructions.
- Wear UV‑blocking eyewear for the operator and anyone in view of the device; do not remove during treatment.
- Protect untreated skin with SPF 30+ sunscreen as directed; avoid treating over active eruptions or burns.
- For very long sessions, use ~50% duty cycle (e.g., 10 min on / 10 min off); allow surfaces to cool if uncomfortable.
- Keep people and pets out of the treatment area; keep device dry; do not use with damaged cord/plug; device must be properly grounded.
- DANGER – ULTRAVIOLET RADIATION: Overexposure can injure eyes/skin. Some medications/cosmetics increase photosensitivity; consult your physician.
4. Delivery, Site & Setup
- Inspect on delivery; report any damage within 24 hours.
- Site: Near a grounded outlet; away from water/traffic; protect from children.
- Assembly: Attach feet with thumb screws; verify they face correctly; tighten all screws. Insert the “D”‑shaped power plug; connect to grounded outlet.
- Lamp check: Verify prescribed lamp type (UVA, UVA‑1, NB‑UVB, BB‑UVB, Blue).
5. Positioning & Treatment Distance
- 20‑lamp units: Treat at 0 in (on acrylic) or 9 in (23 cm). Select the distance on‑screen before starting.
- 10‑lamp units: Treat from 9 in (23 cm) only.
- Large areas: Stand 9 in from lamps (mark floor to help). Avoid standing closer to prevent streaking/burning.
- Small areas (20‑lamp only): Place the area gently on the acrylic near center; avoid excess pressure.
6. Unlocking & Common Functions
- Tap the blank screen → Daavlin logo appears.
- Tap the logo → enter 7 → press Enter.
- At treatment end, controller beeps and shows delivered dose & elapsed time.
- If power is lost mid‑treatment, ClearLink remembers elapsed time and allows resuming after power is restored.
7. Operating – Guided Mode
- Unlock. Scheduled dose appears.
- Answer: “Were you in pain or red after your last treatment?” Choosing Still red/pain locks the device until one session is skipped.
- (20‑lamp only) Select 0 in or 9 in.
- Review confirmation screen (dose & estimated time). Put on goggles and position yourself.
- Press Enter to start (3‑second countdown). Use Repeat to expose another area before leaving the screen. Pause/Play as needed.
8. Operating – Dosimetric Mode
- Unlock → Treat Patient.
- (20‑lamp only) Choose 0 in or 9 in.
- Set prescribed dose (mJ for UVB, J for UVA as programmed). The integrating dosimeter auto‑adjusts time for lamp/temperature variations.
- Confirm dose & estimated time → goggles on → Enter to start. Use Repeat, Pause/Play as needed.
9. Operating – Timed Mode
- Unlock → Treat Patient.
- (20‑lamp only) Choose 0 in or 9 in.
- Enter time (mm:ss). Tip: use 1:00, not 0:60.
- Confirm time → goggles on → Enter to start. Use Repeat, Pause/Play as needed.
Need percentage increases? Use Daavlin’s Treatment Time Calculator to convert prior time + % increase to a new mm:ss value quickly.
10. FlexRx® Prescriptions
When enabled, remaining exposures show at wake. Refill when 5–10 remain.
- Wake controller → unlock.
- When prompted (or via Prescription), enter the refill code provided by your physician using the on‑screen arrows.
- Press Enter to upload. The controller displays exposures added.
11. Cleaning & Maintenance
Routine
- Dust unit & lamps monthly; deep clean behind acrylic/reflectors annually.
- Replace lamps approximately: UVB ~300 hrs; UVA/Blue ~500 hrs (environment dependent).
Cleaning / Disinfection
- General: Non‑abrasive cloth + mild soap (e.g., Dawn). Avoid paper towels on reflective surfaces.
- Low‑level: Monk™ wipes (per label) or 70% Isopropyl Alcohol with 3‑minute contact time.
- High‑level (between different patients): Use an FDA‑cleared high‑level disinfectant such as Steris Resert XL‑HLD per manufacturer’s instructions.
12. Lamp Replacement
- Unplug device; remove feet; lift top cover.
- Remove clear plastic lamp clasps; press red socket button; lift lamp out with two hands.
- Insert new lamps (same brand as original) until an audible snap; replace all lamps as a set.
- Reassemble; reset lamp hours to 0 (contact support for steps).
13. Calibration & Accuracy (Guided/Dosimetric)
- Calibrate annually or every 100 hours and when lamps change.
- Integrating dosimeter accuracy target: ±5%.
- 20‑lamp units have two output values on the certificate (0 in and 9 in).
14. Operating Specifications
- Ambient Temperature: 15–30°C (59–86°F)
- Relative Humidity: 10–95% (non‑condensing)
- Liquid Ingress: IPX0 (no water ingress protection)
- Ocular Hazard Distance: 3 m (9.84 ft)
- Ambient Luminance: 250–500 lux
15. Warranty & Support
- Warranty: New equipment 1 year; Remanufactured equipment 90 days; Lamps 90 days.
- Support: USA/Canada 1‑800‑322‑8546 • Overseas 1‑419‑636‑6304 • Email: service@daavlin.com • www.daavlin.com
16. Downloadable Resources
- 4 Series ClearLink™ Operation Manual (PDF) – add your site link
- Treatment Time Calculator
- Treatment Log Sheet (PDF) – optional