1. Device Overview
The Daavlin M Series is a tabletop phototherapy cabinet for hands, feet, and elbows. It can be prescribed with either the ClearLink™ Home-Control (touchscreen) or the Digital Timer keypad controller. Combination UVA/UVB configurations are available; on combination units the front lamps are UVA and the rear lamps are UVB, with the hood always UVA.
2. Indications for Use
Indicated for physician-directed treatment of diagnosed photoresponsive skin disorders such as psoriasis, vitiligo, and atopic dermatitis (eczema) across pediatric through geriatric populations.
3. Warnings & Safety
- Prescription use only. Follow your physician’s protocol.
- Always wear UV-blocking goggles (100% UVA/UVB) for anyone in view of the lamps.
- For PUVA (psoralens + UVA), protect the eyes from sunlight for 24 hours after taking the drug.
- Danger — Ultraviolet Radiation: Overexposure can injure eyes/skin.
- Keep untreated skin protected (clothing or SPF ≥30). Remove jewelry; place hands/feet/elbows flat on the acrylic guard centered over the lamps.
- For long sessions, use a duty cycle (e.g., 10 minutes on / 10 minutes off) to avoid heat buildup.
- Never treat more than once in a 24-hour period unless specifically directed by your physician.
4. Setup & Assembly
- Select a site: Dry, child-safe area with a grounded outlet; keep away from water.
- Unpack: Base and hood may ship separately. Place the base, then seat the hood (arched openings forward) and connect the hood power cord to the base.
- Power: Insert the “D”-shaped plug into the rear receptacle, then plug into the wall outlet.
- Lamp check: Verify prescribed lamps (e.g., PLL-36W series) are installed per your configuration (UVA, Narrowband UVB, Broadband UVB).
- Hood lamps: Use the hood ON/OFF rocker as desired; on combo units the hood is UVA and does not operate during a UVB treatment.
5. Operating Instructions
5A. ClearLink™ Controller (Guided / Dosimetric / Timed)
Unlocking: Tap the blank screen → tap the Daavlin logo → enter 7 → press Enter. The device self-locks after idle time.
Guided Mode (auto-progress with safety checks)
- After unlocking, the scheduled dose appears. The system asks if you had pain/redness after the last treatment (including a “still red/pain” option that locks the device until one session is skipped).
- Confirm the treatment and estimated time. Put on goggles and position the treatment area; press Enter to begin (3-second countdown).
- Use Pause/Play to hold/resume; press Cancel after pausing to end. Use Repeat to treat additional areas before leaving the treatment screen.
- Integrating dosimetry automatically compensates for lamp output changes due to temperature/aging.
Dosimetric Mode (dose-driven)
- Press Treat Patient, set the prescribed dose with ▲/▼ (mJ for UVB or J for UVA as programmed), confirm, and press Enter to start.
- Pause/Resume/Cancel and Repeat work as above; integrating dosimetry adjusts time to deliver the set dose.
Timed Mode (time-driven)
- Press Treat Patient, set the time with ▲/▼, confirm, and press Enter to start. Tip: enter 1:00 instead of “0:60.”
- Use Pause/Resume/Cancel and Repeat as needed.
End of treatment: The unit beeps and shows delivered dose and elapsed time. If power is lost mid-treatment, ClearLink tracks progress and can resume once power is restored.
5B. Digital Timer Controller (keypad)
Unlocking: Press any key → display shows C0dE → press ▲ seven times to show 0007 → press ENTER.
Set a Treatment Time
- After unlocking, display reads 00:00. Use ▲ to change the flashing digit; use ◄ to move left. Press ENTER to lock the time (e.g., 01:10).
- Press ENTER to begin (indicator lights during treatment). Three beeps signal completion and the final time flashes.
- Special functions: Pause with END or ENTER (display “HOLD”); resume with ENTER; end with END twice. Max time is 59:59. If power fails, the controller recalls elapsed time and resumes on restore.
- Repeat last time: Use the controller’s recall sequence (if equipped) to quickly repeat the previous setting.
- Lamp hours: With lamps off, use the controller’s lamp-hour view sequence to check age.
6. FlexRx™ Prescriptions
- ClearLink: On wake, the FlexRx screen shows remaining exposures. When 5–10 remain, obtain a refill code; upload via the on-screen Prescription flow.
- Digital Timer: When exposures are depleted the display shows a refill prompt (e.g., “PASS”). Enter the 4-digit refill code and press ENTER to load the new total.
7. Cleaning & Maintenance
- Between treatments: Disinfect patient-contact surfaces (use non-abrasive wipes or 70% isopropyl alcohol with adequate contact time). Avoid paper towels on acrylic/reflectors.
- General: Dust monthly; annually remove acrylic to clean lamps/reflectors.
- Calibration / Output checks: ClearLink guided/dosimetric units—recalibrate/verify periodically (e.g., annually or every ~100 hours) and whenever lamps are replaced. Digital Timer—measure output ~every 100 hours and after lamp changes.
- Lamp life (typical): UVB ~300 hours; UVA ~500 hours (environment dependent). Replace all lamps at once; then reset lamp hours.
- Changing lamps: Unplug, remove hood, slide out acrylic, release plastic clasps, press red socket buttons to lift lamps, then reseat new lamps until they snap; reassemble and reset lamp hours.
8. Troubleshooting
- No power / lamps won’t light: Verify outlet and cord; ensure hood cord is connected; check fuses; confirm correct UVA/UVB selection on combination units.
- Controller beeps / ends early: You may have reached the programmed time/dose or pressed Pause; resume or re-enter settings.
- Error codes (timer units): Startup self-tests may show E-codes; persistent errors or “EE” during treatment indicate sensor/current anomalies—contact service.
9. Warranty & Support
Support: Daavlin Service — USA/Canada 1-800-322-8546 • Overseas 1-419-636-6304 • Email: service@daavlin.com • www.daavlin.com
10. Frequently Asked Questions
- What’s the unlock code? ClearLink uses 7; Digital Timer uses 0007 (▲×7 then ENTER).
- Can I treat more than once per day? No—never more than one treatment within 24 hours unless directed by your physician.
- Missed treatments? In Guided Mode, the controller adjusts per prompts; otherwise follow your physician’s reduction schedule.
- How do I repeat for another area? Use Repeat on ClearLink before leaving the treatment screen, or use the timer’s recall sequence to re-use the last time.