1 Series

Modified on Sun, 21 Sep at 9:24 AM

Daavlin 1 Series – ClearLink Controller

1. Device Overview

The Daavlin 1 Series is a home phototherapy device available with two controller options: ClearLink™ (touchscreen) and Digital Timer (DT). Both support 0 in and 9 in treatment distances and are prescribed and supervised by a physician.

2. Controller Types (ClearLink vs. DT)

FeatureClearLink™Digital Timer (DT)
ModesGuided, Dosimetric, TimedTimed
UnlockTap blank screen → tap logo → enter 7 → EnterPower on → display shows C0dE → set 0007 with ▲ and ◀ → Play/Pause
Dose ControlGuided auto-steps and erythema checks; Dosimetric in J/mJ with built‑in sensorTime entry (mm:ss) with character-by-character edit
FlexRx®Upload refill prescription codes on-screenDevice displays Cxyz; obtain 4‑digit refill code from provider

3. Indications for Use

For the treatment of diagnosed skin disorders such as psoriasis, vitiligo, and atopic dermatitis (eczema) under physician direction. Suitable from pediatric through geriatric populations.

4. Warnings & Safety

  • Prescription Use Only. Treatments must be supervised by a licensed physician.
  • Always wear UV‑blocking goggles for anyone in view of the device.
  • Protect untreated skin (e.g., sunscreen SPF 30+ on areas not being treated).
  • Do not treat over burns/eruptions or if experiencing unresolved erythema.
  • Typical thermal duty cycle: ~10 minutes on / 10 minutes off for very long sessions.
  • Keep people and pets out of the treatment area. Avoid water and damaged cords.

5. Setup & Assembly

  1. Site: Dry, ventilated area; near grounded outlet; away from traffic/water.
  2. Unpack: Remove device and accessories. Connect the “D”-shaped power plug, then plug into a grounded outlet.
  3. Optional: Install swivel stand per its instructions (if purchased).
  4. Verify lamps: Confirm lamp type matches prescription (UVA, UVA‑1, NB‑UVB, BB‑UVB, Blue).

6. Positioning & Treatment Distance

Select one of two distances before treating:

  • 0 in (on acrylic): Higher intensity; used for hands/feet resting on the guard.
  • 9 in (23 cm): For larger areas with the device upright and body 9 in away.

Use the distance consistently with the prescription; switching distances changes treatment time.

8. Operating – Digital Timer (DT)

  1. Unlock with key code (see next section).
  2. Display shows 00:00 with rightmost digit flashing.
  3. Use to change the flashing digit; use to move left.
  4. Press Play/Pause to lock in the time, then press again to start (pause/resume with the same key).
  5. After completion, the controller flashes the last treatment in Standby; press Play/Pause twice to repeat.

Example: To enter 01:23 → ▲×3 → ◀ then ▲×2 → ◀ then ▲×1 → Play/Pause → Play/Pause.

9. Unlock Codes & Security

ClearLink™

  1. Tap blank screen → tap Daavlin logo.
  2. Enter 7 → press Enter.

Digital Timer (DT)

  1. Press any button to power on.
  2. When C0dE appears, set 0007 with (seven presses) and Play/Pause.

10. FlexRx® Prescriptions

ClearLink™ – Uploading a Refill

  1. Wake screen → unlock.
  2. When prompted (or via Prescription), enter the refill code provided by your physician.
  3. Confirm; the controller displays exposures uploaded.

Tip: A reminder shows remaining exposures on wake. Refill when 5–10 remain.

DT – Viewing Device Code & Refill

  1. Unlock → press Cancel + + Play/Pause together → from “— — — 1” press four times to “— — — 5” → Play/Pause.
  2. Display shows next Cxyz Device Code (also appears when exposures <20 or exhausted).
  3. Give Cxyz to your physician → Daavlin generates a 4‑digit refill code; enter to restore exposures.

11. Cleaning, Maintenance & Lamps

Routine Care

  • Clean patient-contact surfaces between each treatment.
  • Dust unit and lamps monthly; clean behind acrylic and reflectors annually.
  • Replace lamps: UVB ~300 hrs; UVA/Blue ~500 hrs (environment dependent).

Cleaning & Disinfection

  • General: Non‑abrasive cloth + mild soap (no paper towels on reflective surfaces).
  • Low‑level: Monk™ wipes (per label) or 70% Isopropyl Alcohol (3 min contact).
  • High‑level (between different patients): Use an FDA‑cleared high‑level disinfectant (e.g., Resert XL‑HLD) per manufacturer.

Lamp Replacement (summary)

  1. Unplug. Remove two side screws; lift top cover.
  2. Remove clear plastic clasps; press red socket button; lift lamp out with two hands.
  3. Seat new lamps firmly (audible snap). Replace all lamps as a set; then reset lamp hours to 0.

Calibration (ClearLink Guided/Dosimetric)

Maintain calibration annually or every 100 hours and when lamps change; the integrating dosimeter targets ±5% timer accuracy and ±10% sensor accuracy.

12. Troubleshooting

  • Lamps won't ignite: Confirm unlock, mode/time set, goggles on, and power; if still failing, contact support.
  • Unexpected lock (“Still red/pain”): Skip one treatment to clear (Guided mode safety).
  • Power loss mid‑treatment: ClearLink remembers elapsed time (lamps remain off until resumed). DT does not; restart per prescription.
  • Overheating/discomfort: Use 10‑on/10‑off duty cycle for long sessions; allow cool‑down.
  • FlexRx exhausted: Follow refill steps in Section 10.

13. Operating Specifications

  • Ambient Temperature: 15–30°C (59–86°F)
  • Relative Humidity: 10–95% (non‑condensing)
  • Liquid Ingress: IPX0 (no water ingress protection)
  • Ocular Hazard Distance: 3 m (9.84 ft)
  • Ambient Luminance: 250–500 lux

14. Warranty & Support

  • Device: 1 year; Remanufactured equipment: 90 days; Lamps: 90 days.
  • Support: USA/Canada 1‑800‑322‑8546 • Overseas 1‑419‑636‑6304 • Email: service@daavlin.comwww.daavlin.com