1. Device Overview
The Daavlin 1 Series is a home phototherapy device available with two controller options: ClearLink™ (touchscreen) and Digital Timer (DT). Both support 0 in and 9 in treatment distances and are prescribed and supervised by a physician.
2. Controller Types (ClearLink vs. DT)
Feature | ClearLink™ | Digital Timer (DT) |
---|---|---|
Modes | Guided, Dosimetric, Timed | Timed |
Unlock | Tap blank screen → tap logo → enter 7 → Enter | Power on → display shows C0dE → set 0007 with ▲ and ◀ → Play/Pause |
Dose Control | Guided auto-steps and erythema checks; Dosimetric in J/mJ with built‑in sensor | Time entry (mm:ss) with character-by-character edit |
FlexRx® | Upload refill prescription codes on-screen | Device displays Cxyz ; obtain 4‑digit refill code from provider |
3. Indications for Use
For the treatment of diagnosed skin disorders such as psoriasis, vitiligo, and atopic dermatitis (eczema) under physician direction. Suitable from pediatric through geriatric populations.
4. Warnings & Safety
- Prescription Use Only. Treatments must be supervised by a licensed physician.
- Always wear UV‑blocking goggles for anyone in view of the device.
- Protect untreated skin (e.g., sunscreen SPF 30+ on areas not being treated).
- Do not treat over burns/eruptions or if experiencing unresolved erythema.
- Typical thermal duty cycle: ~10 minutes on / 10 minutes off for very long sessions.
- Keep people and pets out of the treatment area. Avoid water and damaged cords.
5. Setup & Assembly
- Site: Dry, ventilated area; near grounded outlet; away from traffic/water.
- Unpack: Remove device and accessories. Connect the “D”-shaped power plug, then plug into a grounded outlet.
- Optional: Install swivel stand per its instructions (if purchased).
- Verify lamps: Confirm lamp type matches prescription (UVA, UVA‑1, NB‑UVB, BB‑UVB, Blue).
6. Positioning & Treatment Distance
Select one of two distances before treating:
- 0 in (on acrylic): Higher intensity; used for hands/feet resting on the guard.
- 9 in (23 cm): For larger areas with the device upright and body 9 in away.
Use the distance consistently with the prescription; switching distances changes treatment time.
7. Operating – ClearLink™
Guided Mode
- Unlock device. The scheduled dose appears.
- Answer “Were you in pain or red after your last treatment?” (Still red/pain locks the device until one session is skipped.)
- Select distance: 0 in or 9 in.
- Confirm dose & estimated time, put on goggles, position yourself, then Enter to start (3‑sec countdown).
- Use Repeat to treat another area with the same dose before leaving the screen.
Dosimetric Mode
- Unlock → Treat Patient → choose 0 in or 9 in.
- Set prescribed dose (mJ for UVB; J for UVA as programmed).
- Confirm, wear goggles, position, then Enter to begin. Integrating dosimetry auto‑adjusts time.
- Pause/Play as needed; Repeat to expose another area before exiting.
Timed Mode
- Unlock → Treat Patient → set time (mm:ss). Use 1:00 instead of 0:60.
- Confirm, wear goggles, select distance, then Enter to begin.
- Tip: For % increases, use Daavlin’s Treatment Time Calculator (see Downloads).
8. Operating – Digital Timer (DT)
- Unlock with key code (see next section).
- Display shows
00:00
with rightmost digit flashing. - Use ▲ to change the flashing digit; use ◀ to move left.
- Press Play/Pause to lock in the time, then press again to start (pause/resume with the same key).
- After completion, the controller flashes the last treatment in Standby; press Play/Pause twice to repeat.
Example: To enter 01:23 → ▲×3 → ◀ then ▲×2 → ◀ then ▲×1 → Play/Pause → Play/Pause.
9. Unlock Codes & Security
ClearLink™
- Tap blank screen → tap Daavlin logo.
- Enter 7 → press Enter.
Digital Timer (DT)
- Press any button to power on.
- When
C0dE
appears, set 0007 with ▲ (seven presses) and Play/Pause.
10. FlexRx® Prescriptions
ClearLink™ – Uploading a Refill
- Wake screen → unlock.
- When prompted (or via Prescription), enter the refill code provided by your physician.
- Confirm; the controller displays exposures uploaded.
Tip: A reminder shows remaining exposures on wake. Refill when 5–10 remain.
DT – Viewing Device Code & Refill
- Unlock → press Cancel + ◀ + Play/Pause together → from “— — — 1” press ▲ four times to “— — — 5” → Play/Pause.
- Display shows next Cxyz Device Code (also appears when exposures <20 or exhausted).
- Give Cxyz to your physician → Daavlin generates a 4‑digit refill code; enter to restore exposures.
11. Cleaning, Maintenance & Lamps
Routine Care
- Clean patient-contact surfaces between each treatment.
- Dust unit and lamps monthly; clean behind acrylic and reflectors annually.
- Replace lamps: UVB ~300 hrs; UVA/Blue ~500 hrs (environment dependent).
Cleaning & Disinfection
- General: Non‑abrasive cloth + mild soap (no paper towels on reflective surfaces).
- Low‑level: Monk™ wipes (per label) or 70% Isopropyl Alcohol (3 min contact).
- High‑level (between different patients): Use an FDA‑cleared high‑level disinfectant (e.g., Resert XL‑HLD) per manufacturer.
Lamp Replacement (summary)
- Unplug. Remove two side screws; lift top cover.
- Remove clear plastic clasps; press red socket button; lift lamp out with two hands.
- Seat new lamps firmly (audible snap). Replace all lamps as a set; then reset lamp hours to 0.
Calibration (ClearLink Guided/Dosimetric)
Maintain calibration annually or every 100 hours and when lamps change; the integrating dosimeter targets ±5% timer accuracy and ±10% sensor accuracy.
12. Troubleshooting
- Lamps won't ignite: Confirm unlock, mode/time set, goggles on, and power; if still failing, contact support.
- Unexpected lock (“Still red/pain”): Skip one treatment to clear (Guided mode safety).
- Power loss mid‑treatment: ClearLink remembers elapsed time (lamps remain off until resumed). DT does not; restart per prescription.
- Overheating/discomfort: Use 10‑on/10‑off duty cycle for long sessions; allow cool‑down.
- FlexRx exhausted: Follow refill steps in Section 10.
13. Operating Specifications
- Ambient Temperature: 15–30°C (59–86°F)
- Relative Humidity: 10–95% (non‑condensing)
- Liquid Ingress: IPX0 (no water ingress protection)
- Ocular Hazard Distance: 3 m (9.84 ft)
- Ambient Luminance: 250–500 lux
14. Warranty & Support
- Device: 1 year; Remanufactured equipment: 90 days; Lamps: 90 days.
- Support: USA/Canada 1‑800‑322‑8546 • Overseas 1‑419‑636‑6304 • Email: service@daavlin.com • www.daavlin.com